Products (Note 1)
|
Standard, Spring, Pneumatic and Feather Touch
|
Spring Push Axial Cable
|
AX/5/S
|
Spring Push Radial Cable
|
AXR/5/S
|
Spring Push Axial Cable Feather Touch
|
AT/5/S
|
Spring Push Radial Cable Feather Touch
|
ATR/5/S
|
Pneumatic Axial Cable
|
AX/5/P
|
Pneumatic Radial Cable
|
AXR/5/P
|
Pneumatic Axial Cable Feather Touch
|
AT/5/P
|
Pneumatic Radial Cable Feather Touch
|
ATR/5/P
|
Pneumatic Axial Cable Jet
|
AJ/5/P
|
Pneumatic Radial Cable Jet
|
AJR/5/P
|
Measurement Performance
|
Measurement Range (mm)
|
±5
|
Linearity (% of Reading)
|
(Note 2)
|
0.50
|
Linearity (µm)
|
(Note 2)
|
5
|
Maximum Repeatability (µm)
|
(Note 3)
|
0.15
|
Typical Repeatability (µm)
|
(Note 4)
|
0.07
|
Resolution (µm)
|
(Note 5)
|
|
Pre Travel (mm)
|
(Note 6)
|
0.15
|
Post Travel (mm)
|
(Note 7)
|
0.85
|
Pre Travel adjustment range (mm) (Spring Push Only)
|
1.50
|
Tip Force (N)
|
(Note 8)
|
|
Spring Push ±20 %
|
0.70
|
Spring Push Feather Touch ±20 %
|
0.30
|
Pneumatic ±20 % at 0.4 bar
|
0.70
|
Pneumatic ±20 % at 1 bar
|
2.60
|
Pneumatic Feather Touch ±30 % at 0.4 bar
|
0.18
|
Pneumatic Feather Touch ±30 % at 1 bar
|
1.10
|
Ultra Feather Touch at 0.4 bar
|
N/A
|
Pneumatic Jet ±30 % at 1 bar
|
0.85
|
Temperature Coefficient %FS/°C
|
0.01
|
Electrical Interface (Note 9)
|
LVDT Sensitivity - Plugged ±0.5% (mV/V/mm)
|
40
|
LVDT Sensitivity - Unplugged ±5% (mV/V/mm)
|
105
|
Half Bridge Sensitivity - Plugged ±0.5% (mV/V/mm)
|
14.7
|
Half Bridge Sensitivity - Unplugged ±5% (mV/V/mm)
|
51
|
LVDT Energising Current ±5% (mA/V)
|
2
|
Half Bridge Energising Current ±5% (mA/V)
|
1.2
|
Environmental
|
IP65 with gaiter or IP50 without gaiter
|
-20 to +80
|
+5 to +80 with gaiter or -10 to +80 without gaiter
|
cycles depending on application, typical 10 million in most applications
|
Materials
|
Stainless Steel
|
Nylon, Ruby, Silicon Nitride, Tungsten Carbide
|
Fluoroelastomer or Silicon
|
steel braided or armoured options available
|
Note 6: Distance from the fully out position to start of calibrated measuring range
|
Note 7: Distance from the end of the calibrated measuring range to the fully in position
|
Note 8: Tip force is at the mid point of the measuring range
|
Note 9: LVDT probes are calibrated at 3 V, 5 kHz into a 10 kΩ load (100 kΩ unplugged). Half Bridge at 3 V, 10 kHz into a 2 kΩ load (1 kΩ unplugged). The probes will operate at energising voltages between 1 and 10 V and frequencies between 2 and 20 kHz, but the performance is not specified
|