Gas Measured
|
Hydrogen Cyanide (HCN)
|
Cartridge Part Number
|
MIDAS-S-HCN 1 year standard warranty
MIDAS-E-HCN 2 year extended warranty
|
Sensor Technology
|
3 electrode electrochemical cell
|
Measuring Range (ppm)
|
HCN 0 – 20ppm
|
Minimum Alarm 1 Set Point
|
2.4ppm
|
Repeatability
|
< ± 2% of measured value
|
Linearity
|
< ± 4% of measured value
|
Response Time t
|
< 15 seconds
|
Sensor Cartridge Life Expectancy
|
≥ 24 months under typical application conditions
|
Operating Temperature
Effect of Temperature
Zero
Sensitivity
|
0°C to +40°C (32°F to 104°F)
< ± 0.008ppm / °C (0°C to 20°C)
< ± 0.03ppm / °C (20°C to 40°C)
< ± 2.5% of measured value / °C
|
Operating Humidity (continuous) Effect of Humidity
Zero
Sensitivity
|
15 – 90% rH
TBA
< ± 1% of measured value / % rH
|
Operating Pressure
|
90 – 110kPa
|
Effect of Position
|
No effect in typical application
|
Long Term Drift
Zero
Sensitivity
|
No effect
< ± 2% of measured value / month
|
Calibration Gas
|
Hydrogen Cyanide (HCN)
|
Challenge Gas (Bump Test)
|
Sulphur Dioxide (SO)
|
Warm Up Time
|
< 10 minutes
|
Storage Temperature
|
﹢5°C to +25°C (+41°F to +77°F)
|